Description
1. Commercial presentation
The absorbable multifilament polyglactin 910 (PGLA) suture, reference P1L18RHRP3630, is the choice of the professional seeking an absorbable thread with predictable behaviour, excellent handling and maximum knot security. It combines a 1/2 circle cylindrical body needle HRP 36 mm with a Trocar Ultra Plus point, a USP 3/0 gauge and a violet 70 cm thread with high visibility in the surgical field. Ideal for soft tissue approximation, ligatures and ophthalmic surgery. Presentation in a box of 12 units, ready for use in the operating theatre.
2. Technical Explanation
Polyglactin 910 is a synthetic copolymer composed of 90 % glycolide and 10 % L-lactide, structured as a coated braided multifilament that ensures smooth passage through tissue, excellent knotting and high initial tensile strength. Its absorption occurs by hydrolysis between 56 and 72 days, with progressive and predictable loss of strength, no need for later removal and minimal tissue reaction.
The 1/2 circle cylindrical body needle with Trocar Ultra Plus point (designation HRP36) penetrates tissues with less resistance and without cutting effect on adjacent tissue, minimising trauma and improving control in delicate areas such as the ocular surface and lax tissues. The product is supplied sterile by ethylene oxide (EO).
3. Technical Specifications
Main features of reference P1L18RHRP3630:
| Characteristic | Specification |
|---|---|
| Reference | P1L18RHRP3630 |
| Material | Polyglactin 910 (PGLA): 90% glycolide / 10% L-lactide |
| Thread structure | Coated braided multifilament |
| USP / metric gauge | 3/0 (metric 2) |
| Thread length | 70 cm |
| Colour | Violet |
| Needle shape | 1/2 circle |
| Needle body | Cylindrical |
| Needle size | 36 mm (HRP36) |
| Needle tip | Trocar Ultra Plus (Taperpoint type) |
| Sterilisation | Ethylene oxide (EO) |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
Absorption profile and tensile strength retention
| Period | Tensile strength retention |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 30% of initial tensile strength |
| Total absorption | 56 – 72 days by hydrolysis |
4. Equivalences and Cross-References
Polyglactin 910 is one of the classic compositions in synthetic absorbable multifilament suture, so this Prime reference is clinically equivalent to the Vicryl™ from Ethicon (Johnson \u0026 Johnson) lines, to the Safil from B.Braun range, to Polysorb from Medtronic/Covidien and to the polyglactin 910 ranges from SMI (Surgicryl), both in chemical composition and in absorption profile (56-72 days) and tensile strength retention. The HRP needle with Trocar Ultra Plus point is functionally equivalent to the reinforced conical needles with penetration point (taperpoint type) used by the main manufacturers.
5. Selection tips and variants of interest
The USP 3/0 gauge with HRP 36 mm needle and Trocar Ultra Plus point is especially suitable for delicate tissues, ophthalmic surgery and precision ligatures. If you prefer a longer thread, the 90 cm variant is available (P1L13RHRP3630); for HRP needle of 30 mm or 26 mm, see P1L18RHRP3030 (3/0, 70 cm) and P1L18RHRP2630 (3/0, 70 cm). For smaller or larger gauges with the same HRP36 needle profile, consider P1L18RHRP3620 (2/0), P1L18RHRP3610 (USP 0) and P1L13RHRP3640 (4/0, 90 cm). There are also undyed versions in the family (P1L28RHRP3601 and P1L28RHRP3620). Single-use material: store in a cool, dry place, protected from sunlight and inside its sterile packaging until use. Consult our technical-commercial team for special conditions by volume and hospital supply contracts.