Description
1. Introduction
The Nonin N-7500FO Tabletop Pulse Oximeter (reference N-7500FO) is the professional reference solution for continuous monitoring of oxygen saturation (SpO₂) and pulse rate in Magnetic Resonance Imaging (MRI) environments. Designed by Nonin Medical, world leader in oximetry, this tabletop monitor allows the clinical team to monitor the patient in real time without compromising safety or the quality of the diagnostic image.
Thanks to its non-conductive fibre optic technology, it eliminates the risks associated with the use of conventional metal cables inside the examination room, being suitable for fixed and mobile MRI installations. It is an essential tool for radiology services, anaesthesia, critical care units, hybrid operating theatres and biomedical research centres. It is supplied without a sensor: compatible fibre optic sensors are purchased separately depending on the patient type and the required cable length.
2. Technical explanation of the equipment
The Nonin N-7500FO is a tabletop pulse oximeter with bidirectional fibre optic architecture. The display unit is installed outside the magnetic field (control room) and is connected via non-conductive fibre optic cables to the optical sensors placed on the patient inside the MRI room, passing through the magnetic isolation tunnel. This completely eliminates radiofrequency (RF) induced currents, avoiding RF burns and image artefacts.
Internally it uses Nonin’s PureSAT® PureLight technology, which applies intelligent pulse-by-pulse filtering to provide accurate measurements even in complex clinical conditions, with movement or low perfusion. The device features a large LED screen with 3-digit numeric display in green, pulse quality indicator (amber), pulse intensity bar graph, sensor failure indicator, visual and audible SpO₂ and pulse alarms with configurable thresholds and internal memory of up to 70 hours of data in continuous use. It is also compatible with the Nonin nVision data management and monitoring system, allowing oximetry data to be integrated into clinical and research workflows.
During use inside the MRI room, the device must operate exclusively on internal battery; connection to the mains is reserved for recharging and for use outside the MRI environment. Always read the operator’s manual for installation requirements and advice.
3. Technical Specifications
| Parameter | Value |
| Manufacturer | Nonin Medical |
| Model / Reference | N-7500FO |
| Set type | Tabletop pulse oximeter for MRI (without sensor) |
| Measured parameters | SpO₂ and pulse rate |
| SpO₂ display range | 0–100 % |
| SpO₂ accuracy | ±2 digits (70–100 %) with fibre optic sensor |
| Pulse rate range | 18–321 ppm |
| Pulse rate accuracy | ±3 digits without movement (18–300 ppm); ±3 digits in low perfusion (40–240 ppm) |
| Display | Large LED: 3 digits, 7 segments, in green |
| Indicators | Pulse quality, sensor failure, pulse intensity, alarm, low battery |
| Alarms | Adjustable visual and audible alarms for SpO₂ and pulse (75 dBA high / 64 dBA low) |
| Internal memory | Up to 70 hours of continuous use |
| Sensor compatibility | Nonin bidirectional fibre optic sensors: N-8000FC-20, N-8000FC-30, N-8000FI-20, N-8000FI-30 |
| System compatibility | Nonin nVision monitoring system |
| Application | Adults, paediatric and infant in MRI environment |
| Includes sensor | No (sold without sensor) |
| Power supply and conditions | Value |
| Mains power | 100–240 V AC, 50–60 Hz |
| DC input | 12 V DC, 1.5 A (in MRI only with battery) |
| Internal battery | NiMH 7.2 V |
| Battery life | Minimum 12 hours with full charge |
| Recharge time | Maximum 4 hours |
| Dimensions | Approx. 219 (W) x 92 (H) x 142 (D) mm |
| Weight | Approx. 900 g with battery |
| Enclosure protection | IPX2 |
| Electrical classification | Applied part Type BF, continuous operation |
| Operating temperature | 0 °C to 40 °C |
| Measurement wavelengths | Red 660 nm / Infrared 910 nm |
| Warranty | 3 years |
Compatible fibre optic sensors (sold separately)
| Reference | Description | Cable length |
| N-8000FC-20 | Adult/paediatric fibre optic sensor | 6 metres |
| N-8000FC-30 | Adult/paediatric fibre optic sensor | 9 metres |
| N-8000FI-20 | Infant/paediatric fibre optic sensor | 6 metres |
| N-8000FI-30 | Infant/paediatric fibre optic sensor | 9 metres |
4. Equivalences and Cross-References
The N-7500FO is the specific version for magnetic resonance imaging of the Nonin tabletop oximeter family, which includes the standard model N-7500 (with conventional sensor) for general use outside the MRI environment. The FO variant is sold without sensor and requires exclusively Nonin bidirectional fibre optic sensors from the 8000F series (N-8000FC-20, N-8000FC-30 for adult/paediatric and N-8000FI-20, N-8000FI-30 for infant/paediatric), verified in the official Nonin catalogue. The monitor is also compatible with the Nonin data management system nVision.
Pulse oximeters specifically for the magnetic resonance imaging environment are a highly specialised niche: most generalist oximetry manufacturers do not offer an equivalent non-conductive fibre optic tabletop unit. Below is the alternative from another manufacturer, cross-checked against the official technical data sheet and authorised distributors; none is interchangeable 1:1 with the N-7500FO without prior technical validation by the MRI service.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| EMCO Meditek | OXI-MRI 3600 (MRI-compatible pulse oximeter) | OXI-MRI 3600 | Comparable by category (not interchangeable 1:1) | Non-magnetic fibre optic sensor of 20 feet (≈6 m) valid for 0.5–3 T environments; LED/LCD screen readable from the control room; integrated rechargeable battery (~6 h, approx. 8 scans); includes sensor, unlike the N-7500FO which is supplied without sensor (source: official EMCO Meditek technical data sheet and authorised distributors). |
In the market, manufacturers operating with MRI-compatible oximetry include Masimo (systems such as Radical-7 with MRI-compatible connectivity and patient cables, SET® technology) and Medtronic / Nellcor (MRI-compatible extended sensors and cables with OxiMax™ technology), as well as MIPM (oximetry sensors for MRI) and BERRY (SpO₂ sensors compatible with MRI), in addition to the OXY-MRI module from BIOPAC aimed at physiological research. In all cases, field-by-field verification (sensor type, cable length, compatibility with the scanner and classification) must be carried out against the manufacturer’s official technical data sheet before any replacement.
There is no manufacturer-validated equivalence table for a 1:1 cross-reference between the N-7500FO and systems from other manufacturers; consult the technical department before using in tenders.
5. Selection tips and variants of interest
- Select the sensor according to the patient: use the N-8000FC series sensors for adult and paediatric patients, and the N-8000FI series for infant patients.
- Choose the appropriate cable length: 6 metres (N-8000FC-20 / N-8000FI-20) for control rooms close to the scanner, or 9 metres (N-8000FC-30 / N-8000FI-30) if the distance to the magnetic isolation tunnel is greater.
- This equipment is supplied without a sensor: remember to add the compatible fibre optic sensor(s) to the order for immediate start-up.
- MRI installation: the display unit must be located in the control room and, during the scan, the equipment must operate in battery mode; read the operator’s manual for installation requirements.
- Manage the alarms: configure the SpO₂ and pulse thresholds according to the centre’s protocol; the system includes patient safety mode and customisable default values.
- Variants of interest: if your service also needs monitoring outside the MRI environment, consider the Nonin N-7500 model with conventional sensor, as well as the range of fibre optic sensors from the 8000F series and Nonin accessories available in our catalogue.